| Model Number |
ZA9003 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Section a2, a4,and a5: unknown, asked but not available.Section d6a: na as the lens was not implanted.Section d6b: na as the lens was not implanted; therefore, not explanted section h3-other (81): the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record and complaint history for production order for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the monofocal non preloaded intraocular lens haptic was bent.There was no patient injury.Surgeon implanted lens from another manufacturer.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d-9: device available for evaluation: yes.Section d-9: device date returned to manufacturer: 08-aug-2023.Section h-3: device evaluated by manufacturer: yes.Device evaluation: product evaluation was performed under magnification.The complaint lens was received cut in half and both haptics were bent.The lens was cleaned and no additional issues were identified.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order was performed.The search revealed that no other complaints were received for this purchase order.Conclusion: the complaint issue reported could not be confirmed and no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|