• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JJGC S.A. TI TITAMAX EX IMPL TI (4.1)4.0X11; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JJGC S.A. TI TITAMAX EX IMPL TI (4.1)4.0X11; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 109.589
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the removal of a dental implant during surgery without the replacement of another dental implant is a known inherent risk of the procedure due to either lack of primary stability of the implant (patienor procedure related).It may also include the removal of an implant after osseointegration due to either the clinician's or patient¿s decision.The manufacturer¿s trend analysis confirms that usually procedural errors and/or patient's condition contribute to the event.
 
Event or Problem Description
The clinician reports that the day the implant was placed in fdi 22, failure occurred upon insertion.Details of surgery: primary stability not achieved and implant surface not completely covered with bone.Patient presented with inadequate bone quality/quantity.The device was forwarded to the manufacturer.There were no reported patient injuries or complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI TITAMAX EX IMPL TI (4.1)4.0X11
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
Manufacturer (Section G)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR   81270-200
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key17501578
Report Number3008261720-2023-016673
Device Sequence Number6392279
Product Code DZE
UDI-Device Identifier07898237568416
UDI-Public07898237568416
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K101207
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/29/2021
Device Catalogue Number109.589
Device Lot Number86866
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/10/2023
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient SexFemale
-
-