Catalog Number 1012580-16
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Difficult to Advance (2920); Activation Failure (3270)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented with a free perforation in the right coronary artery.The 3.5x16mm graftmaster covered stent was attempted to be advanced; however, failed to cross due to anatomy.A 2.8x19mm graftmaster was advanced and deployed successfully; however, the perforation was larger than expected.A 2.8x16mm graftmaster covered stent was advanced to the perforation; however, during inflation the device would not inflate even at 20 atmospheres.It was noted the graftmaster met resistance with anatomy during advancement.Therefore, an additional graftmasters 3.5x19mm was attempted to be advanced; however, failed to cross due to anatomy.A 2.8x16mm graftmaster was used to successfully complete the procedure and seal the perforation.There was no adverse patient sequela and no clinically significant in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.H6: medical device problem code 2017 - above rbp.The additional graftmaster devices referenced in b5 are filed under separate medwatch report numbers.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.It was reported that the graftmaster was inflated to a maximum pressure of 20 atmospheres (atm).It should be noted in the graftmaster rapid exchange (rx), coronary stent graft system, domestic, instructions for use specifies the rated burst pressure (rbp) is 16 atm and clearly states not to exceed the rbp.It is unknown if the exceeded rated burst pressure contributed to the reported event.A conclusive cause for the reported difficulties could not be determined; however, the subsequent treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Common Device Name
CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17501827
Report Number2024168-2023-08640
Device Sequence Number6513916
Product Code MAF
UDI-Device Identifier08717648176340
UDI-Public08717648176340
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Catalogue Number1012580-16
Device Lot Number2062041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/20/2023
Supplement Date Received by Manufacturer08/23/2023
Initial Report FDA Received Date08/10/2023
Supplement Report FDA Received Date09/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight68 KG
-
-