| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2; -6.50/+1.0/75 (sphere/cylinder/axis) implantable collamer lens into the patient's right eye (od) on (b)(6) 2023.The patient experienced excessive vaulting, elevated intraocular pressure without pupil block, significant reduction of idiro-corneal angles, additionally medical intervention of combigan and cosopt 3 times a day.Lens remains implanted.The problem is not resolved.The surgeon noted there is a planned lens exchange.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - health effect clinical code: 4581 - significant reduction of irido-corneal angles.H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - lens explanted and replaced in a different surgery on (b)(6) 2023 with a shorter length lens and problem was resolved.Clear cornea, formed ac, icl in place.Claim# (b)(4).
|
| |
|