Model Number VTICM5_13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Event or Problem Description
The reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2; -6.50/+1.0/75 (sphere/cylinder/axis) implantable collamer lens into the patient's right eye (od) on (b)(6) 2023.The patient experienced excessive vaulting, elevated intraocular pressure without pupil block, significant reduction of idiro-corneal angles, additionally medical intervention of combigan and cosopt 3 times a day.Lens remains implanted.The problem is not resolved.The surgeon noted there is a planned lens exchange.
 
Additional Manufacturer Narrative
H6 - health effect clinical code: 4581 - significant reduction of irido-corneal angles.H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
 
Additional Manufacturer Narrative
B5 - lens explanted and replaced in a different surgery on (b)(6) 2023 with a shorter length lens and problem was resolved.Clear cornea, formed ac, icl in place.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia,, CA 91016
6269301410
MDR Report Key17502541
Report Number2023826-2023-03460
Device Sequence Number6466990
Product Code QCB
UDI-Device Identifier00840311311118
UDI-Public00840311311118
Combination Product (Y/N)N
Initial Reporter CountryRP
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberVTICM5_13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2023
Supplement Date Received by Manufacturer09/05/2023
Initial Report FDA Received Date08/10/2023
Supplement Report FDA Received Date10/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL-SFC-45, LOT#-UNK.; FOAM TIP PLUNGER MODEL-FTP, LOT#-UNK.; INJECTOR MODEL-MSI-PF, LOT#-UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-