• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMBOLD FIBERED DETACHABLE COIL SYSTEM; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMBOLD FIBERED DETACHABLE COIL SYSTEM; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83910
Medical Device Problem Codes Material Protrusion/Extrusion (2979); Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the coil was protruding from the target location.An embold fibered 10x50 detachable coil system was selected for use in the embolization procedure.The coil was intended to be deployed distally in the varicocele vein.The physician advanced the coil to the target location of deployment and ensured that the couplers were outside of the microcatheter before breaking the proximal release perforation to deploy the coil, but the coil did not detach.The physician retracted the pull wire further, but the coil still did not deploy.The microcatheter was pulled back and an alternate pusher wire was utilized to push the coil into the target location.Due to the additional manipulation, the proximal tip of the coil was deployed more proximally in the vessel than intended.The proximal portion of the coil protruded from the target lesion.No further action was taken as a result of the unintended coil placement.The procedure was successfully completed and there were no reported patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBOLD FIBERED DETACHABLE COIL SYSTEM
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17504645
Report Number2124215-2023-41371
Device Sequence Number14977786
Product Code KRD
UDI-Device Identifier08714729983651
UDI-Public08714729983651
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/14/2026
Device Model Number83910
Device Catalogue Number83910
Device Lot Number0031616598
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/14/2023
Initial Report FDA Received Date08/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-