|
This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation and device evaluation.The device was evaluated.And the reported issue (balloon failure to deflate) could not be confirmed, (the balloon could not be deflated for testing).However, the complete device inspection results are as follows: minor kinks were observed, on the insertion tube portion of the sheath.The fluoro tip was examined under the microscope.And the balloon was confirmed, to have popped.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event (balloon failure to deflate) could not be determined.However, the issue was likely, caused by the following mechanism: 1) a force beyond the strength resistance was applied to the balloon, when retrieving stones such as calculi, pancreatic and common bile duct stones.2) the fixed area at the rear end of the balloon broke.And it was displaced to the distal end.3) a hole for inflating and deflating the balloon was exposed.4) the balloon could not deflate.The event can be detected/prevented, by following the instructions for use (ifu), which state: ¿before use, prepare and inspect the instrument as instructed below.Should any irregularity be observed, do not use the instrument.Use a spare instead.Damage or irregularity may compromise patient or user safety, pose an infection-control risk, cause tissue irritation, perforation, bleeding or mucous membrane damage.And may result in more severe equipment damage¿.¿inflate the balloon only with air.Inflation with anything other than air may hinder expansion and contraction of the balloon¿.¿if it is difficult to pull the plunger of the premeasured syringe when deflating the balloon, detach the premeasured syringe from the air-feeding port¿.¿if resistance is too strong and withdrawal is difficult, adjust the angle of the endoscope until the instrument can be withdrawn smoothly.Forcible withdrawal of the instrument could damage the instrument and/or endoscope¿.¿when using an endoscope equipped with a forceps elevator, do not withdraw the instrument from the endoscope, if the forceps elevator is up.This could damage the instrument¿.This supplemental report includes a correction to d4 (lot number), e2, e3, and h8 from the initial medwatch.Also, an update has been made to d9 and h3 from the initial medwatch.Olympus will continue to monitor field performance for this device.
|