Manufacturer#s ref# (b)(4).Technical investigation concluded: a dhr review have been completed.No deviation or changes were found during manufacturing of the device.Trended investigation concluded: the connector breaks, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.Clinical assessmnet concluded: it has been reported that the charger overheated and melted at plug port.It was plugged into the wall outlet and sitting on a nightstand.There was no harm to the user, and no mention of property damage.Original accessories were used.The chargers have been tested according to applicable safety standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.In case of moisture, sweat or dirt in the connector, there can be high temperatures inside the connector due to power dissipation when connected to the charger plug.There can be melting of the plastic around the connector and is visible to the naked eye.It is highly unlikely that the user would touch in case there appears to be melting or high temperatures.In the unlikely event of the user touching this overheated device, it can lead to superficial or minor injury that would in most cases require no medical intervention.There are no new clinical aspects (e.G., unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation risk assessment concluded: no new hazardous situations revealed and no need to update the risk register.This is a combined initial and final report.
|