• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. IPC® FOOTPEDAL - LEGEND EHS; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. IPC® FOOTPEDAL - LEGEND EHS; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE Back to Search Results
Model Number EF201RF
Medical Device Problem Code Device Remains Activated (1525)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that handpiece activated without pedal being pushed, there is no patient impact.
 
Additional Manufacturer Narrative
Product analysis of device is in progress, analysis results of the device will be submitted in supplemental report.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
H3: product analysis found that there was oxidation inside the connector.H6: codes are updated.Previously applied fdm b21, fdr c21, fdc d16 and img g0408201 codes no longer applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IPC® FOOTPEDAL - LEGEND EHS
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key17510443
Report Number1045254-2023-00625
Device Sequence Number5535943
Product Code HBE
UDI-Device Identifier00643169931503
UDI-Public00643169931503
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K081475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEF201RF
Device Catalogue NumberEF201RF
Device Lot Number223208999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2023
Supplement Date Received by Manufacturer09/24/2023
Initial Report FDA Received Date08/11/2023
Supplement Report FDA Received Date10/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/28/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-