|
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found that while checking the distal end of the device under a microscope, the probe was detached (the missing portion was not returned).The device was attached to the usg-400 ultrasonic generator / esg-400 electrosurgical high frequency generator while a probe check was performed and the device failed the probe check with an error code of u509 (probe damage error code), and the ptfe pad (teflon pad) was found to be partially lifted but no metal was exposed.The investigation is ongoing, a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation, as well as additional information/clarification to b5.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the probe broke due to contact with a surgical instrument or the non-insulated area of grasping section.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1.During the output activation in seal & cut mode, the probe came into contact with hard tissue, metal objects or surgical instruments.This resulted in scratches.2.A force to activate the output in seal & cut mode, or a force to grasp the body tissue was applied to the probe.Therefore, cracks were branching from the scratches on the probe.3.The probe broke when that when removing the tb-0009ofx from the td-tb400, a load might have been applied.Or 1.Grasping section was closed without grasping anything while the output was activated in seal & cut mode (this includes after tissue resection).This caused the tissue pad to wear out.2.The tissue pad was worn out.This has resulted in the non-insulated area and the distal end of the probe coming into contact with each other.3.The output was activated in seal & cut mode in state of above description.This has resulted in scratches (contact marks) on the two distal ends of the probe and on the grasping section.4.A force was applied to the probe to activate the output in the seal & cut mode, or a force was applied to grasp the tissue.This resulted in cracks branching at the scratched area.5.The probe broke when that when removing the tb-0009ofx from the td-tb400, a load might have been applied.The following are the instructions for use which state: "do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.Make sure that the thunderbeat instrument and the transducer are connected firmly.If they are secured erroneously or only with the hand, energy output may not be available and damage to the transducer or excessive heating of the exterior may also result.Even if output is available, the functionality and durability may be compromised.Use the provided torque wrench and stabilizer for the connection and disconnection and be sure to follow the instructions given in this manual.If another tool, a hand, or excessive force is used for securing, incomplete connection or damage may result to the thunderbeat instrument, or the transducer resulting in the impossibility of disconnection.If proper connection is impossible, there may be a deformation such as crushing or bending in some parts.Inspect the instrument well and immediately stop using it if any irregularity is observed.Stop application of force to the torque wrench when an audible click is heard.Otherwise, the system may not work properly and serious injury to the surgeon, surgical staff, and/or the patient may result".Olympus will continue to monitor field performance for this device.
|