DEPUY ORTHOPAEDICS INC US SIGMA HP UNI TIB TRAY SZ3 LMRL; EARLY INTERVENTION : KNEE TIBIAL TRAY Back to Search Results
Catalog Number 102451300
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the surgery, it was found that the insert could not be locked and it was easy to loosen and pull out.Replacing the insert caused that the procedure delayed by approximately 30 minutes.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: could not lock the insert.During the surgery, it was found that the insert could not be locked.And it was easy to loosen and pull out.After replacing another insert with the same lot number.The loosening situation happened again.Please check the detail in the video which is in the attachment.The doctor decided to continue to complete the surgery, but it may cause potential patient risk.Replacing the insert caused that the procedure delayed by approximately 30 minutes.The patient is stable now.The product was not returned to depuy synthes, however a video was provided for review.See attachment "(b)(4)".The video investigation was able to confirm the loosening condition of the uni insert, however with the evidence provided nothing indicative of a device nonconformance was found for the sigma hp uni tib tray sz3 lmrl or if it contributed with the loosening condition.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was unconfirmed as the observed condition of the [sigma hp uni tib tray sz3 lmrl] would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing record evaluation was performed for the finished device jg8000 lot number, and no nonconformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI TIB TRAY SZ3 LMRL
Common Device Name
EARLY INTERVENTION : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key17511813
Report Number1818910-2023-16614
Device Sequence Number12146422
Product Code NPJ
UDI-Device Identifier10603295002338
UDI-Public10603295002338
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K070267
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number102451300
Device Lot NumberJG8000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date08/11/2023
Supplement Report FDA Received Date08/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-