This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device was missing components, fell apart and had cosmetic damage.The device also failed pretests for visual assessment.The assignable root cause was traced to component failure.Udi: (b)(4).
|
|
It was reported that during an unspecified surgical procedure it was observed that the attachment device would not lock onto the motor device.During in-house engineering evaluation, it was determined that the device was missing components, fell apart and had cosmetic damage.The device also failed pretests for visual assessment.It was unknown if there was a delay to the surgical procedure.A spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|