K2M, INC. 751 MILLER DRIVE SE, LEESBURG, VIRGINIA K2M SPINE SCREWS; PEDICLE SCREW SPINAL SYSTEM, ADOLESCENT IDIOPATHIC SCOLIOSIS
|
Back to Search Results |
|
| Medical Device Problem Code |
Therapeutic or Diagnostic Output Failure (3023)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/19/2012
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Empire medical; during a mast lumbar fusion the surgeon used a medtronic navlock awl and tap, and then attempted to use a different spine companies screws (k2m).The screws would not thread after attempting multiple times.Therefore, the surgeon became frustrated with the spine rep (non medtronic) and aborted navigation.A c-arm was brought in and used the other spine company's instruments to successfully finish the surgery without navigation.We all agreed that the medtronic tap was not compatible (different threads) with the spine company (k2m).(b)(4).This report reflects information received by fda in the form of a notification per 803.22 (b)(2).
|
| |
|
Search Alerts/Recalls
|
|
|