| Model Number |
CLH-SC |
| Medical Device Problem Code |
No Visual Prompts/Feedback (4021)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation, and the customer's allegation was confirmed and was reproduced.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that the halogen light source does not turn on, (no light), and there was a disconnection issue.There were no reports of patient harm or impact associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to consumption of the lamp.Olympus will continue to monitor field performance for this device.
|
| |
|