| Model Number |
NEB KIT W ADAPTER STERILE WATER 1000ML |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: 81 other ¿ at this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not returned yet.
|
| |
|
Event or Problem Description
|
|
It was reported that the vyaire medical that the neb kit w adapter sterile water 1000ml is attached to oxygen flow and placed on a patient, the water would leak out of the entrainment port all over the floor.The customer confirmed that there was no patient harm associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: g3, g6, h2, h3, h6 and h10.Result of investigation: based on the investigation and per physical samples received, we cannot confirm the reported defect since the sample was functionally inspected per pqas ck0005 etal performing a vacuum test as well as gage rings diameter verification and no issues were found.Additionally, the device history records of the fg kit prefill neb w neb cap sterile water part number ck0010 with lot number 0004250078 was reviewed and no issues were found.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|