• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS HEALTHCARE (SUZHOU) CO., LTD. INGENUITY CT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS HEALTHCARE (SUZHOU) CO., LTD. INGENUITY CT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number INGENUITY FLEX
Medical Device Problem Codes Unintended System Motion (1430); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: (b)(4).
 
Event or Problem Description
This complaint has been evaluated based on the information provided.The customer reported that after a patient scanning was completed, the patient couch suddenly dropped while lowering the couch.Philips field service engineer(fse) inspected system and identified that ball bearing and ball screw were defective.Fse had the new ball bearing service kit installed and returned the system to customer for clinical use.Patient was frightened but was not actually harmed.A similar product regarding couch dropping downwards rapidly was recently assessed as potential for serious injury if it were to recur.Based on that assessment, we are considering this event reportable out of abundance of caution.Philips has started an investigation of this complaint.
 
Additional Manufacturer Narrative
The customer, (b)(6), reported that after a patient scanning was completed, the patient couch suddenly dropped while lowering the couch.The system was in clinical use during the event.A philips field service engineer (fse) went onsite to evaluate the reported issue and confirmed the patient was not injured and did not receive any treatment.A system inspection determined that the vertical screw rod of the patient couch was not damaged, the cause of the failure was the long-term wear of the bearing.Fse replaced the bearings service kit, and the system was returned to use in good working order.In conclusion, a ball bearing failure was the root cause of the event.A review of the risk management file indicates the problem reported by the customer is of low potential severity, which would not reasonably cause or contribute to death or serious injury if the problem were to reoccur.Therefore, based on the investigation¿s conclusion, this issue has been determined not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENUITY CT
Common Device Name
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS HEALTHCARE (SUZHOU) CO., LTD.
no. 258, zhongyaun road
suzhou industrial park
suzhou jiangsu 21502 4
CH  215024
Manufacturer (Section G)
PHILIPS HEALTHCARE (SUZHOU) CO., LTD.
no. 258, zhongyaun road
suzhou industrial park
suzhou jiangsu 21502 4
CH   215024
Manufacturer Contact
estelle hilas
100 park avenue, suite 300
orange village, OH 44122
MDR Report Key17538212
Report Number3009529630-2023-00007
Device Sequence Number6333694
Product Code JAK
UDI-Device Identifier00884838059528
UDI-Public00884838059528
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
160743
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberINGENUITY FLEX
Device Catalogue Number728317
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/21/2023
Supplement Date Received by Manufacturer07/21/2023
Initial Report FDA Received Date08/13/2023
Supplement Report FDA Received Date03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-