| Model Number |
ALFA01 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During procedure, the operator wanted to grasp the jj catheter after the isiris alfa was introduction into the bladder.The claw of the grasper detached.A second endoscope was then used to retrive the detached claw from the bladder cavity.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported failure (detached claw) was verified from returned sample.Based on investigation results it is concluded that the grasper claws was activated without having visual control and accidentally hit the jj catheter during grasping which caused it to bend outwards (opening over the opening range) and thereby breaking within the patient.
|
| |
|