• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD CARIBE LTD. BD¿ SPINAL NEEDLE WITH QUINCKE BEVEL; ANESTHESIA CONDUCTION NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD CARIBE LTD. BD¿ SPINAL NEEDLE WITH QUINCKE BEVEL; ANESTHESIA CONDUCTION NEEDLE Back to Search Results
Catalog Number 405164
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that medication leaked from the bd¿ spinal needle with quincke bevel and syringe connection site during use.The following information was provided by the initial reporter, translated from spanish: "when the syringe is connected to the needle, the medication leaks and much of it is lost, the syringe is not well attached to the needle, which causes much of the medication to be lost.".
 
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that medication leaked from the bd¿ spinal needle with quincke bevel and syringe connection site during use.The following information was provided by the initial reporter, translated from spanish: "when the syringe is connected to the needle, the medication leaks and much of it is lost, the syringe is not well attached to the needle, which causes much of the medication to be lost.".
 
Additional Manufacturer Narrative
Investigation summary: one video received by our quality team for investigation.Through visual evaluation, it is observed the process of a puncture is being performed, during this process the syringe leaks through the upper part of the hub.A device history review was performed for lot 2108321, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this issue.Final products in this manufacturing line, for this reference are sampled and they are subjected to visual and functional inspections during the different manufacturing sub-processes according to procedures.Based on the quality team's investigation, we are not able to identify a root cause related to our manufacturing process at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ SPINAL NEEDLE WITH QUINCKE BEVEL
Common Device Name
ANESTHESIA CONDUCTION NEEDLE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos IA
Manufacturer (Section G)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos IA
Manufacturer Contact
jennifer suh
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key17541744
Report Number2618282-2023-00065
Device Sequence Number7570384
Product Code BSP
UDI-Device Identifier00382904051648
UDI-Public00382904051648
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K210978
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number405164
Device Lot Number2108321
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2023
Supplement Date Received by Manufacturer08/31/2023
Initial Report FDA Received Date08/14/2023
Supplement Report FDA Received Date09/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-