Catalog Number 191128
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 08/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Event or Problem Description
A fresenius field service technician (fst) reported that the blood pump rotor cover on the 2008t machine was cracked and tore through the fresenius combi set bloodlines during a patient¿s hemodialysis (hd) treatment, causing a blood leak to occur.Upon follow up with the user facility registered nurse (rn), it was confirmed that the event occurred during the patient¿s treatment and the machine did provide air detected alarms to alert the staff to the issue.The rn stated that the issue was discovered when it was visually observed by the medical staff.There was no defect or damage noted to the bloodlines prior to the occurrence of this issue.The rn stated that the bloodlines appeared to have been torn by the machine.The patient¿s estimated blood loss (ebl) was approximately 100 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient successfully completed treatment on a different machine with new supplies.The blood pump rotor was replaced to resolve the reported issue.The machine has been returned to service.The complaint device is available to be returned for physical evaluation by the manufacturer.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.However, an on-site evaluation was performed by fresenius field service technician (fst).The blood pump rotor was replaced to resolve the reported issue.A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was able to confirm the reported failure mode.During the machine inspection, the fst reported the blood pump rotor cover on the 2008t machine was cracked and tore through the fresenius combi set bloodlines during a patient¿s hemodialysis (hd) treatment, causing a blood leak to occur.Therefore, the complaint event was confirmed.
 
Event or Problem Description
A fresenius field service technician (fst) reported that the blood pump rotor cover on the 2008t machine was cracked and tore through the fresenius combi set bloodlines during a patient¿s hemodialysis (hd) treatment, causing a blood leak to occur.Upon follow up with the user facility registered nurse (rn), it was confirmed that the event occurred during the patient¿s treatment and the machine did provide air detected alarms to alert the staff to the issue.The rn stated that the issue was discovered when it was visually observed by the medical staff.There was no defect or damage noted to the bloodlines prior to the occurrence of this issue.The rn stated that the bloodlines appeared to have been torn by the machine.The patient¿s estimated blood loss (ebl) was approximately 100 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient successfully completed treatment on a different machine with new supplies.The blood pump rotor was replaced to resolve the reported issue.The machine has been returned to service.The complaint device is available to be returned for physical evaluation by the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. W/O CDX BLUE STAR
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key17541841
Report Number0002937457-2023-01200
Device Sequence Number20466705
Product Code KDI
UDI-Device Identifier00840861102105
UDI-Public00840861102105
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K173972
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number191128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2023
Supplement Date Received by Manufacturer09/07/2023
Initial Report FDA Received Date08/14/2023
Supplement Report FDA Received Date09/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS BLOODLINES; FRESENIUS BLOODLINES; FRESENIUS DIALYZER; FRESENIUS DIALYZER
Patient Age77 YR
Patient SexFemale
Patient Weight75 KG
-
-