• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 ECHELON; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 ECHELON; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 190-5091-150
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: product analysis of 190-5091-150, lotno:b541773 visual inspection/damage location details: the echelon-10 total length was measured to be ~155.4cm.The echelon-10 useable length was measured to be ~147.6cm.Upon visual examination, no issues were found with the echelon-10 hub.No bends or kinks were found with the echelon-10 catheter body.However, saline residue was found on catheter body at ~0.5cm from separated tip.The distal marker band and distal tip were found to be separated and not returned.The tubing material of the catheter separated end exhibited stretching with jagged edges.No other anomalies were observed.Conclusion: based on the device analysis and reported information, the customer¿s report of ¿prod damaged/deformed out of pkg¿ was confirmed; however, the broken end of the catheter exhibited plastic deformation (stretching and jagged edges) which indicate that the catheter separated when exceeding the tensile strength of the tubing material.In this event, user error likely contributed to the during the removal of the echelon-10 micro catheter tip from within the tip protector and then separated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that an echelon 10 catheter tip had a stitch on it.While preparation of the catheter (flushing the catheter), there was a little stitch coming out of the catheter.The catheter was flushed as per instructions for use (ifu).The devices were prepared as per ifu.There was no patient involved with this event.There was no damage or evidence of tampering to device packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key17542094
Report Number2029214-2023-01372
Device Sequence Number20466567
Product Code KRA
UDI-Device Identifier00847536006294
UDI-Public00847536006294
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K042187
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/30/2025
Device Model Number190-5091-150
Device Catalogue Number190-5091-150
Device Lot NumberB541773
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2023
Initial Report FDA Received Date08/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-