SYNTHES GMBH 4.2MM TROCAR 210MM; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.010.070
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery on (b)(6), 2023, the frn 10x400 l gt was inserted into a heavier patient.When inserting the nail, there was resistance.Once the nail was fully inserted, green triple sleeves were inserted through the jig and malleted.The inner trocar ended up bending, so it was taken out and then drilled.The drill bit hit the nail, so a new triple sleeve was inserted.The drill bit hit the nail again.The nail was removed, and a tfna was inserted instead with no issues.The surgery was successfully completed without any delay.There was no patient consequence.This report involves one 4.2mm trocar 210mm.This is report 2 of 5 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.2MM TROCAR 210MM
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key17542287
Report Number8030965-2023-10267
Device Sequence Number20683265
Product Code FZX
UDI-Device Identifier10886982067425
UDI-Public(01)10886982067425
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.070
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/31/2023
Initial Report FDA Received Date08/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
10 / TI CANN FRN / GT 400 / LEFT - SILE; 4.2MM TROCAR 210MM; UNK - DRILL BITS: TRAUMA; UNK - GUIDES/SLEEVES/AIMING: SLEEVE
Patient SexUnknown
-
-