| Catalog Number |
1574 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Other remarks: n/a.Corrected data: n/a.
|
| |
|
Event or Problem Description
|
|
It was reported that a hair was found in the package upon opening.Therefore, a new unit was used instead.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).Customer complaint cannot be confirmed as no product sample was available to perform a proper investigation and determine the root c ause.There was no sample returned for this complaint.Therefore, a product from a previous complaint was reviewed.No foreign matter found inside the pouches for these samples.In current manufacturing procedure, 100% visual inspection after assembly process is conducted.Thus, any defective products will be culled out during this process.Therefore, it is very unlikely condition at the manuf acturing area that the product having abnormality to be released for shipment.If the alleged defect samples become available at a later date, this complaint will be updated accordingly.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that a hair was found in the package upon opening.Therefore, a new unit was used instead.No patient involvement.
|
| |
|