| Model Number |
M0062101170 |
| Medical Device Problem Codes |
Inflation Problem (1310); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a nephromax nephrostomy balloon was used during a percutaneous nephrolithotomy (pcnl) procedure performed on (b)(6), 2023.During the procedure, the balloon continues to "waste" and would not fully inflate even when inflated to the maximum pressure.The balloon was removed and was attempted to inflate again multiple times after slightly repositioning the wire but sill unsuccessful.After inspection, it was noticed that there was a kink in the balloon which was causing the balloon from fully inflating.The procedure was completed with another nephromax nephrostomy balloon.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Section e: this event was reported by the bsc sales representative.The health care facility is: (b)(6) phone: (b)(6) block h6: imdrf device code a0406 captures the reportable event of balloon irregular shape.
|
| |
|