• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NEPHROMAX; CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NEPHROMAX; CATHETER, NEPHROSTOMY Back to Search Results
Model Number M0062101170
Medical Device Problem Codes Inflation Problem (1310); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a nephromax nephrostomy balloon was used during a percutaneous nephrolithotomy (pcnl) procedure performed on (b)(6), 2023.During the procedure, the balloon continues to "waste" and would not fully inflate even when inflated to the maximum pressure.The balloon was removed and was attempted to inflate again multiple times after slightly repositioning the wire but sill unsuccessful.After inspection, it was noticed that there was a kink in the balloon which was causing the balloon from fully inflating.The procedure was completed with another nephromax nephrostomy balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Section e: this event was reported by the bsc sales representative.The health care facility is: (b)(6) phone: (b)(6) block h6: imdrf device code a0406 captures the reportable event of balloon irregular shape.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEPHROMAX
Common Device Name
CATHETER, NEPHROSTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17546135
Report Number3005099803-2023-04309
Device Sequence Number6534649
Product Code LJE
UDI-Device Identifier08714729012641
UDI-Public08714729012641
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K121614
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2025
Device Model NumberM0062101170
Device Catalogue Number210-117
Device Lot Number0030025641
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/18/2023
Initial Report FDA Received Date08/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1