• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRAEGER MEDICAL SYSTEMS, INC JM-105 JAUNDICE METER; JAUNDICE METER/BILIRUBINOMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRAEGER MEDICAL SYSTEMS, INC JM-105 JAUNDICE METER; JAUNDICE METER/BILIRUBINOMETER Back to Search Results
Catalog Number MU20105
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code Jaundice (2187)
Date of Event 07/13/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow-up report will be submitted upon completion of this investigation.
 
Event or Problem Description
It was reported that the draeger jm105 jaundice meter was providing inconsistent/low readings.No adverse patient impact was reported.
 
Additional Manufacturer Narrative
A complaint was received regarding the draeger jm105 jaundice meter was providing inconsistent/low readings.Additional information was provided; however, the customer provided only two measurement which is not enough to perform data analysis.The response to the clinical questions confirmed the meter was being used correctly.Therefore, the reported low reading could not be verified.The meter was returned to the dreager medical repair center and no malfunction was identified.As a precautionary measure the sensor set was replaced, and the meter was calibrated and returned to the customer.Root cause of the low readings could not be determined.For customer satisfaction, the meter was replaced to resolve the issue.No further issues have been reported.
 
Event or Problem Description
It was reported that the draeger jm105 jaundice meter was providing inconsistent/low readings.No adverse patient impact was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JM-105 JAUNDICE METER
Common Device Name
JAUNDICE METER/BILIRUBINOMETER
Manufacturer (Section D)
DRAEGER MEDICAL SYSTEMS, INC
3135 quarry road
telford PA
Manufacturer (Section G)
DRAEGER MEDICAL SYSTEMS, INC
3135 quarry road
telford PA
Manufacturer Contact
3135 quarry road
telford, PA 
2156602400
MDR Report Key17546187
Report Number2510954-2023-00014
Device Sequence Number6712559
Product Code MQM
UDI-Device Identifier04049098000246
UDI-Public(01)04049098000246(11)210409(93)MU20105-21
Combination Product (Y/N)N
PMA/510(K) Number
K133175
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMU20105
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/02/2023
Supplement Date Received by Manufacturer05/17/2024
Initial Report FDA Received Date08/14/2023
Supplement Report FDA Received Date05/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-