• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SSX-615
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/17/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported right side "extracapsular rupture".Device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Device has been explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, d6b, h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key17546406
Report Number9617229-2023-13234
Device Sequence Number17734831
Product Code FTR
UDI-Device Identifier10888628033429
UDI-Public10888628033429
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/04/2026
Device Catalogue NumberSSX-615
Device Lot Number3521341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/24/2023
Supplement Date Received by Manufacturer09/28/2023
Initial Report FDA Received Date08/14/2023
Supplement Report FDA Received Date10/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight93 KG
-
-