|
|
| Catalog Number |
71174922 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a plate procedure surgery, the tip of one (1) depth gauge for 2.0mm screws broke.The procedure was resumed, without any surgical delay, using a s+n back-up device.Patient was not injured as consequence of this problem.
|
| |
|
Event or Problem Description
|
|
It was reported that during handling activities on a plate procedure surgery, the tip of one (1) depth gauge for 2.0mm screws broke external to patient.The procedure was resumed, without any surgical delay, using a s+n back-up device.Patient was not injured as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event, as the event happened during handling activities.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly, however, as a courtesy the result of investigation are being shared.H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|