• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR CANNULA; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR CANNULA; CAT Back to Search Results
Model Number OPT318
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Fisher and paykel healthcare (f&p) are currently in the process of obtaining further information regarding the reported event.We have also requested the return of the complaint cannula to f&p new zealand for investigation.We will provide a follow up report upon completion of our investigation.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in brazil, that an opt318 optiflow junior nasal cannula had broken.There was no patient consequence.
 
Additional Manufacturer Narrative
(b)(4).Method: the complaint opt318 optiflow junior nasal cannula was received at fisher & paykel healthcare (f&p) new zealand, where it was visually inspected.Results: the visual inspection of the complaint opt318 optiflow junior nasal cannula confirmed that the one of the tubes had detached at the swivel grip joint.Glue was observed at the detached tube end.Conclusion: we are unable to determine the cause of the reported event.However, it is most likely caused by the tube being pulled.All optiflow junior cannula are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The device history record was reviewed and confirmed that the subject opt318 optiflow junior nasal cannula was manufactured according to specification.The user instructions which accompany the opt318 optiflow junior nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: - ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Under excessive load, the cannila maydisconnect to prevent forces being transferred to the patient.- appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death.- do not crush or stretch tube.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility, via a fisher and paykel healthcare (f&p) field representative, in brazil, that an opt318 optiflow junior nasal cannula had broken.There were no reported patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR CANNULA
Common Device Name
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key17548692
Report Number9611451-2023-00752
Device Sequence Number17816169
Product Code CAT
UDI-Device Identifier09420012432308
UDI-Public(01)09420012432308(10)2102530958(11)230227
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOPT318
Device Catalogue NumberOPT318
Device Lot Number2102530958
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2023
Supplement Date Received by Manufacturer09/20/2023
Initial Report FDA Received Date08/15/2023
Supplement Report FDA Received Date10/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-