MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 3501615
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 07/27/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 56-year-old caucasian female patient underwent a primary breast augmentation surgery with implantation of a 150cc mentor smooth round moderate profile saline breast implant prosthesis.Post-operatively, the patient reported she experienced bilaterally deflated breast implants.A consultation with a medical professional has been conducted; however, at the time of this report, mentor has received no information regarding explantation or an expected explantation date.Several follow-ups have been conducted, and no additional information has been received.If more information becomes available, mentor will submit a supplemental report accordingly.Refer to manufacturing report number 1645337-2023-09500 for the contralateral event.
 
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803. this report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: the patient has not undergone explantation or reoperation at this time.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On june 30, 2025, mentor was notified that the patient underwent bilateral breast implant removal without replacements on (b)(6) 2025.Date of explantation: (b)(6) 2025.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On august 05, 2025, the mentor analysis lab received the suspect medical device for evaluation.On august 08, 2025, mentor completed an evaluation on the returned device.Device evaluation: mentor then conducted visual inspection, leak testing, and microscopic examination of the device.Visual analysis of the returned device revealed that the breast implant had a tear on the anterior view, measuring 0.1 cm, approximately.Leak testing was performed in accordance with mentor's procedures.No additional leak sites were detected.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND MODERATE PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key17549071
Report Number1645337-2023-09499
Device Sequence Number17841811
Product Code FWM
UDI-Device Identifier00081317001188
UDI-Public(01)00081317001188(17)260224(10)9688954(21)9688954-003
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/24/2026
Device Catalogue Number3501615
Device Lot Number9688954
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2023
Supplement Date Received by Manufacturer06/30/2025
08/05/2025
Initial Report FDA Received Date08/15/2023
Supplement Report FDA Received Date07/25/2025
08/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age57 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-