ETHICON ENDO-SURGERY, LLC. HS SYNERGY CURVED BLADE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number SNGCB
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 8/15/2023.D4 batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure the blade no longer seemed to be working properly.Upon closer inspection the blade had snapped into 2 pieces.The piece which came off was located and put to one side with the rest of the blade.Second device opened and no further issues occurred for remainder of case.Surgeon did acknowledge that he may have come into contact with bone when using the device, which may have been the cause of the blade breaking.No patient impact (surgeon made no complaint of this and no further issues occurred when second device opened) and only a pause of 2-3 minutes whilst opening second device.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/12/2023.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 10/5/2023.D4 batch #: t9588f.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the distal tip of the blade broken off and it was returned with the device.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.During functional testing on gen11 an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device was disassembled to verify the condition of the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.  once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.  a manufacturing record evaluation was performed for the finished device batch t9588f and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HS SYNERGY CURVED BLADE
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key17549165
Report Number3005075853-2023-05775
Device Sequence Number6683872
Product Code LFL
UDI-Device Identifier10705036002802
UDI-Public10705036002802
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K012176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Catalogue NumberSNGCB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/31/2023
Supplement Date Received by Manufacturer09/24/2023
09/19/2023
Initial Report FDA Received Date08/15/2023
Supplement Report FDA Received Date09/12/2023
10/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR AND HANDPIECE
Patient SexUnknown
-
-