EXACTECH, INC. LOGIC CR TIB INSERT SLOPE+, SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-48-4011 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
07/24/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) pending evaluation (d10) concomitant device(s): (b)(6), 02-010-03-0240 - logic cr femoral cem, left, sz 4.(b)(6), 02-012-45-4040 - lgc tibial fit tray cem sz 4f / 4t.(b)(6), 200-02-35 - three peg patella 35mm.
|
| |
|
Event or Problem Description
|
|
It was reported that this 72 y/o male patient's left knee was revised due to wear.Surgeon performed a poly and patella swap.Patient was last known to be in stable condition following the event.Unable to obtain x-rays.Product not returning; hospital kept.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The revision reported was likely the result of tibial insert and patellar wear.The root cause of the device wear could not be confirmed from the images provided.Additional reasons for the reported revision may be inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and radiographs and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|