| Model Number |
CLV-S45 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned and evaluated, and the customer¿s allegation was not confirmed.The output connector was broken and cannot be connected due to excessive stress to the output socket, and the power cord has fallen off due to mishandling.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified; however, the event most likely occurred due to a worn-out socket.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that the will aim for xenon light source over-illuminated, light control no longer works.The event occurred during preparation for use.There was no patient harm associated with the event.The device was returned and evaluated, and it was found that the output connector is broken and cannot be connected, and the power cord has fallen off, this mdr (medical device report) is being submitted to capture the reportable malfunction found during the device evaluation.
|
| |
|