Catalog Number 261222
Medical Device Problem Code Fail-Safe Did Not Operate (4046)
Health Effect - Clinical Code Ambulation or Postural Difficulties (2544)
Date of Event 07/25/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A facility reported a perforator failed to disengage causing cerebral contusion and opening of the meninge.It was treated with coagulation of the cerebral cortex and intraparenchymal hematoma evacuation.The event led to 10 minutes surgical delay.Patient consequences: proprioceptive ataxia (gait disturbances).
 
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Perforator was returned for evaluation: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the perforator unit was inspected using the unaided eye.The unit was heavily soiled with organic material, it had bone fragments lodged between the inner and outer drill.The label was destroyed.Ifu testing was performed with no observed anomalies.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.The complaint could not be verified through failure analysis.Root cause - the root cause is undetermined and was unable to be confirmed in the complaint evaluation.Product was received for analysis and the investigation could not confirm the complaint.Additional information received: when the patient was discharged on september 5, the doctor reported the following comments: "neurologically we find a coherent and oriented patient, presenting no language disorder.The patient evokes an altered spatial perception, great fatigue, but also concentration disorders.This hemineglect and left hemiparesis persists with a defect grip, but in good recovery thanks to physiotherapy.The patient also mentions coordination problems when carrying out certain tasks.He has no headaches, nor any other sign of intracranial hypertension.There is no comitial crisis." "this patient did not have consultation at our center since.".
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN 11MM DISPOSBL PERFORATR
Common Device Name
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key17549887
Report Number3014334038-2023-00137
Device Sequence Number12504414
Product Code HBF
UDI-Device Identifier10381780513605
UDI-Public10381780513605
Combination Product (Y/N)N
PMA/510(K) Number
K071931
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number261222
Device Lot Number7221965
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2023
Supplement Date Received by Manufacturer09/22/2023
Initial Report FDA Received Date08/15/2023
Supplement Report FDA Received Date10/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-