• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S X-FLOW PROSTATECTOMY CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S X-FLOW PROSTATECTOMY CATHETER; INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL Back to Search Results
Catalog Number AB6318
Medical Device Problem Codes Inflation Problem (1310); Failure to Deflate (4060); Unintended Deflation (4061)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
According to available information, this device required replacement due to difficult inflation.The device was difficult to inflate and impossible to deflate.The device eventually self-deflated.The balloon was intact after device removal.It is suspected that the problem was with the valve.No other adverse patient effects were reported.
 
Additional Manufacturer Narrative
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
After receiving this complaint, we searched for other complaints and found none on the lot 9018842.Checking the quality databases revealed one corrective and preventive action are ongoing documentary investigation was performed about this lot, no issue were registered during production.On august we received one used sample.After disinfection, medical device was tested.The balloon can be inflated with water but the valve is leaking and the green piston inside the valve is askew and seems to be pushed in.On 27th october, we received the investigation from our subcontractor : the cause of the defect is a valve problem (offset piston).
 
Event or Problem Description
According to available information, this device required replacement due to difficult inflation.The device was difficult to inflate and impossible to deflate.The device eventually self-deflated.The balloon was intact after device removal.It is suspected that the problem was with the valve.No other adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-FLOW PROSTATECTOMY CATHETER
Common Device Name
INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
CMF-SARLAT
9 avenue edmond rostand
sarlat-la-caneda
FR  
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key17549895
Report Number9610711-2023-00171
Device Sequence Number15169290
Product Code EZL
UDI-Device Identifier03600040144183
UDI-Public3600040144183
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberAB6318
Device Lot Number9112206_AB63181002
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/15/2023
Supplement Report FDA Received Date01/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-