DEPUY IRELAND - 9616671 PIN TRL +4 NEUT 520DX36ID; HIP INSTRUMENTS : ACETABULAR TRIALS Back to Search Results
Catalog Number 221836452
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the trial liner is scratched/gouged.
 
Additional Manufacturer Narrative
(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4) investigation summary product has scratches against underside and gouges on rim.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from attachment the photo investigation revealed several scratches on the cup surface of the pin trl +4 neut 520dx36id, confirmed the reported allegation.No other issues were observed.This type of damage is consistent with other tools and hard edges coming in contact with the device, properly handling and attention to the approved use of the device diminishes the risk of failure.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed for pin trl +4 neut 520dx36id.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, the potential cause is traced to unintended use error, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Device history lot a date code was provided, which indicates that the device was manufactured on (8/15/2005).A manufacturing records evaluation (mre) was not performed since a valid finished good lot number was not provided for this device.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: product has scratches against underside and gouges on rim.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.The device associated with this report was returned to depuy synthes for evaluation.Visual inspection revealed deep scratches on the convex and concave surface of the pin trl +4 neut 520dx36id.This type of damage is consistent with other tools and hard edges coming in contact with the device, properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the [pin trl +4 neut 520dx36id] would contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the date code (pg0805) provided is not a valid finished goods lot.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: product has scratches against underside and gouges on rim.The product damage documented in this pc has been identified during loan kit inspection, by the loan kit technician.The event date and alert date are the date that the inspection took place.There is no surgeon, procedure, or patient details available.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from attachment "(b)(4)." the photo investigation revealed several scratches on the cup surface of the pin trl +4 neut (b)(6), confirmed the reported allegation.No other issues were observed.This type of damage is consistent with other tools and hard edges coming in contact with the device, properly handling and attention to the approved use of the device diminishes the risk of failure.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed for pin trl +4 neut (b)(6).There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, the potential cause is traced to unintended use error, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Device history lot: a date code was provided, which indicates that the device was manufactured on (8/15/2005).A manufacturing records evaluation (mre) was not performed since a valid finished good lot number was not provided for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN TRL +4 NEUT 520DX36ID
Common Device Name
HIP INSTRUMENTS : ACETABULAR TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17550684
Report Number1818910-2023-16773
Device Sequence Number17858421
Product Code HWT
UDI-Device Identifier10603295101352
UDI-Public10603295101352
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 08/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number221836452
Device Lot NumberPG0805
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2023
Supplement Date Received by Manufacturer08/17/2023
09/21/2023
09/21/2023
09/22/2023
Initial Report FDA Received Date08/15/2023
Supplement Report FDA Received Date08/17/2023
09/21/2023
09/21/2023
09/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/15/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-