| Catalog Number |
SSF-695 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed a broken on radius assessed as an unidentified (tear) opening (shell thickness within spec).Additional observations: ¿ crease fold observed on the device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.The device has been explanted.
|
| |
|