MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK AXS INFINITY LS PLUS .091IN ID X 90CM-US; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK AXS INFINITY LS PLUS .091IN ID X 90CM-US; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number INC-11196-90
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a neurovascular procedure the subject catheter unraveled when pulled out of the patient's body.The procedure was completed successfully.No further information is available.
 
Additional Manufacturer Narrative
H3 other text : the device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during a neurovascular procedure the subject catheter unraveled when pulled out of the patient's body.The procedure was completed successfully.No further information is available.
 
Additional Manufacturer Narrative
Based on the results of the dhr (device history record) review, there is no indication that the device, labeling or packaging failed to meet its specifications when released.The subject device is not available; therefore, functional testing as well as visual testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned to the as reported catheter shaft broken/fractured during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXS INFINITY LS PLUS .091IN ID X 90CM-US
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key17552991
Report Number3008881809-2023-00415
Device Sequence Number4801781
Product Code DQY
UDI-Device Identifier07613327459203
UDI-Public07613327459203
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K152876
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 10/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2024
Device Catalogue NumberINC-11196-90
Device Lot Number18015-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/23/2023
Supplement Date Received by Manufacturer09/28/2023
Initial Report FDA Received Date08/15/2023
Supplement Report FDA Received Date10/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-