| Model Number |
EHB13/14 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that during a dental procedure the user facility was using their 13/14 heidbrink root tip pick to remove the patient's tooth when the tip detached and became stuck in the patient's mouth.The report indicated that the patient will return to have the tip removed.To date, hu-friedy has not received confirmation from the user facility that this has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to hu-friedy for evaluation and it was found that both the right and left tips of the device were broken and showed signs of corrosion.The fractured surface on the right tip showed signs of overload failure.The 13/14 heidbrink root tip pick is a delicate instrument designed for removal of root tips or fragments, not to pry the tooth loose.The fractured surface on the left tip had evidence of corrosion likely from poor cleaning practices.The hu-friedy instrument reprocessing guide states "3.2 inspection inspect all instruments after the cleaning and rinsing step for corrosion, damaged surfaces, and impurities.Do not further use damaged instruments." no additional issues have been reported.
|
| |
|