• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. CLOSURE TOP; POLARIS 5.5 SPINAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. CLOSURE TOP; POLARIS 5.5 SPINAL SYSTEM Back to Search Results
Catalog Number 14-500500
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during final tightening intra-operatively, the threads of a cypher closure top were deformed without any fragments detaching.Another closure top was used to complete the case without patient impacts.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Additional Manufacturer Narrative
Additional information in d4: udi number, h4, and h6: component, investigation type, findings, and conclusions.Device evaluation: the device was not returned and no photos were provided, so an evaluation is unable to be performed.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to off-axis forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that during final tightening intra-operatively, the threads of a cypher closure top were deformed without any fragments detaching.Another closure top was used to complete the case without patient impacts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLOSURE TOP
Common Device Name
POLARIS 5.5 SPINAL SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
sabrina abla
10225 westmoor dr.
westminster, CO 80021
7206965158
MDR Report Key17556879
Report Number3012447612-2023-00236
Device Sequence Number17372727
Product Code NKB
UDI-Device Identifier00880304973879
UDI-Public(01)00880304973879(10)J7103933
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K123549
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number14-500500
Device Lot NumberJ7103933
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/16/2023
Supplement Report FDA Received Date06/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-