| Catalog Number |
3558-2 |
| Medical Device Problem Codes |
Fracture (1260); Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a warehouse discovered a pathfinder reduction screwdriver with a fractured tip after it was returned from a hospital.The hospital did not provide any patient or surgical information with the instrument.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections in d9 and h3.Additional information in d4: udi number, h4, and h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed that the tip of the device was fractured, and the laser markings were faded and difficult to read.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that a warehouse discovered a pathfinder reduction screwdriver with a fractured tip after it was returned from a hospital.The hospital did not provide any patient or surgical information with the instrument.
|
| |
|