| Catalog Number |
7209485 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, the loop retriever of a set meniscus mender was soon disassembled between head and shaft when it was used; however it was not necessary to be retrieved from the patient.Surgery resumed after a non-significant delay of 5 minutes; with the same device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).H10 h3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not sealed in its original packaging.The returned open box confirms the device and batch.The meniscus mender set is in the plastic tray and both of the suture threaders are fractured between the hub and the shaft.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the device drawings found a stress relief collar is now attached at the connection between the shaft and the hub of the suture capture loop.The root cause was associated with device design.A design change was implemented as a result of a corrective action initiated to address the reported issue.No further containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|