| Model Number |
18625 |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: other, other text: the device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.E1.Initial reporter phone number exceeded allowable field length, initial reporter phone number is as follows: (b)(6) h3.Product not returned.
|
| |
|
Event or Problem Description
|
|
The customer reported the "emma doesn't measure correctly" as the readings were too high.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.Initial reporter phone number exceeded allowable field length, initial reporter phone number is as follows: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported the "emma doesn't measure correctly" as the readings were too high.No patient impact or consequences were reported.
|
| |
|