• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon deflation issue occurred.The target lesion was located in the arteriovenous fistula.A 4.0 x 80, 75cm mustang balloon catheter was advanced for dilatation.However, during the procedure with a saline-contrast ratio of 3:7, it was observed that the balloon was inflated at 17 atmospheres and did not deflate perfectly.The balloon was completely removed in a deflated state through the sheath and the procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: mustang 4.0 x 80, 75cm, was received for analysis.A visual examination of the returned device found that the balloon was tightly wrapped.There is no evidence to suggest the balloon had been inflated.An examination of the balloon material identified no issues.A visual examination observed no damage to the tip of the device.A visual and tactile examination found no kinks or damage to the shaft of the device.Both markerbands were undamaged and present in the correct position on the shaft of the device.This concludes the product analysis.
 
Event or Problem Description
It was reported that balloon deflation issue occurred.The target lesion was located in the arteriovenous fistula.A 4.0 x 80, 75cm mustang balloon catheter was advanced for dilatation.However, during the procedure with a saline-contrast ratio of 3:7, it was observed that the balloon was inflated at 17 atmospheres and did not deflate perfectly.The balloon was completely removed in a deflated state through the sheath and the procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key17558852
Report Number2124215-2023-39919
Device Sequence Number4879914
Product Code FGE
UDI-Device Identifier08714729794363
UDI-Public08714729794363
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K103751
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2023
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0025961184
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/24/2023
Supplement Date Received by Manufacturer08/22/2023
Initial Report FDA Received Date08/16/2023
Supplement Report FDA Received Date09/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-