| Model Number |
421571 |
| Medical Device Problem Code |
Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported " the catheter tip is sealed with sealing part of package.".
|
| |
|
Additional Manufacturer Narrative
|
|
E.1: complainant city: (b)(6).Complainant state/province: (b)(6).Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092.Manufacturing site: 3005778470.
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|
Additional Manufacturer Narrative
|
|
The pictures related to the complaint were provided and evaluated.The defect reported was confirmed.The issue on the product is detectable during production.100% visual inspection with focus on the issue is carried out.Lot # 2a03369 was packed in february 2022 on packaging machine p006, total lot amount (b)(4).Review of the dhr lot 2a03369 showed that all relevant tests required during the manufacturing and packaging process and final product release had been fulfilled.No nonconformity related to the issue reported had been registered during the packaging process of the mentioned lot.During complaint review, the increased number of complaints related to the issue - catheter squeezed in welding packaging, has been observed.Capa tw (b)(4) has been open to investigate the issue.Within the capa as immediate action operators were retrained on visual inspection that is focused on the reported defect.Was determined that cpm slightly increased, but the incidence of defect occurrence is not high.Lot in question was produced before the retraining was carried out.No harm was reported.The defect is visible and was detected before use.The corrective action implemented previously, that covered implementation of guides under ccr-dmr-00059, did not ensure 100% help to hold catheters on the right place in the cavities.Only reduce potential risk of catheters squeezing in weld.No another complaint of this nature has been received on the lot.No additional action is required, and this complaint will be closed.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 3005778470.
|
| |
|
Search Alerts/Recalls
|