AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEM FOR ADULTS, DOUBLE SWIVEL ELBOW; VAP CLOSED SUCTION CATHETERS & ACCESSORIES Back to Search Results
Model Number 22106-5
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Codes Airway Obstruction (1699); Low Oxygen Saturation (2477)
Date of Event 05/11/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the closed suction catheter had an extra piece of plastic occluding the trach care, which prevented the closed suction catheter from moving freely through and potentially occluding the airway.The product was in use for five minutes.The patient desaturated and deteriorated.The patient was manually bagged and saturations stabilized.A new device was used without any further incident.The patient was reported to be ok.Per additional information received on (b)(6) 2023, ¿staff described it [location of blockage] as the trach care barrel - there was a thin film of plastic occluding the trach care barrel, which also prevented the suction catheter from moving freely through, and potentially occluding the airway.¿ a new trach care was placed with the same lot number, and no issues occurred with the new device.The deterioration was witnessed by staff and the patient recovered after immediate medical attention.The patient remains in intensive therapy unit (itu) ¿ stable.
 
Additional Manufacturer Narrative
The product involved in the report has been returned and the investigation remains in progress at this time.A review of the device history record is in-progress.Manufacturing site updated; all information previously submitted under 9611594-2023-00075 has been transferred to 8030647-2023-00109.All information reasonably known as of 14 aug 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
H6: investigation findings appropriate term/code not available: blockage identified the device history record for lot 30124923 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.Based on sample evaluation, there is a rigid plastic at the base of the female swivel.The plastic obstructs the majority of the double swivel elbow (dse).Advancement of the catheter through the double swivel elbow (dse) was not possible.The plastic is not loose, but appears to be part of the component.The incident was confirmed as reported.The root cause is manufacturing-related.All information reasonably known as of 16 oct 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CLOSED SUCTION SYSTEM FOR ADULTS, DOUBLE SWIVEL ELBOW
Common Device Name
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key17564665
Report Number8030647-2023-00109
Device Sequence Number12435966
Product Code BSY
UDI-Device Identifier00609038983165
UDI-Public00609038983165
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2026
Device Model Number22106-5
Device Catalogue NumberN/A
Device Lot Number30124923
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2023
Supplement Date Received by Manufacturer10/10/2023
Initial Report FDA Received Date08/17/2023
Supplement Report FDA Received Date10/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age70 YR
Patient SexMale
Patient Weight57 KG
-
-