| Catalog Number |
130003 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: date of event is unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the battery case was broken.Patient involvement is unknown.
|
| |
|
Event or Problem Description
|
|
Additional information received via email.The event occurred during testing.Event date was updated from unknown to 02-march-2023.The was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: h6.Health effects; updated.G1,2 email is: regulatory.Responses@icumed.Com.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was returned for analysis.Visual inspection confirmed the customer's complaint.Battery compartment is damaged.However, this is not due to case damage but because the screw that holds it in place was missing.The root cause for this event is the impact on device.The product is beyond a year from manufacture date and there was no indication of a manufacturing defect during the investigation, so a device history record review was not performed.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.D.4.Full udi number: (b)(4).
|
| |
|