• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAYLIS MEDICAL COMPANY INC. VERSACROSS CONNECT LAAC ACCESS SOLUTION; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAYLIS MEDICAL COMPANY INC. VERSACROSS CONNECT LAAC ACCESS SOLUTION; DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION Back to Search Results
Lot Number VMFC290523
Medical Device Problem Codes Difficult to Remove (1528); Defective Device (2588); Component or Accessory Incompatibility (2897)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported that during a watchman procedure to treat left atrial appendage closure (laac), a versacross connect kit was selected for use.Prior to transseptal puncture, the mechanical j-wire was in superior vena cava (svc) and it frayed/ split when it was pulled into versacross connect sheath.There was resistance felt when withdrawing the wire from the introducer.Both the versacross connect and double curve sheath were removed from the body.Hence to resolve the issue, a new kit was opened and the procedure was completed successfully.No patient complications reported.The device is expected to be returned for analysis.
 
Event or Problem Description
It was reported that during a watchman procedure to treat left atrial appendage closure (laac), a versacross connect kit was selected for use.Prior to transseptal puncture, the mechanical j-wire was in superior vena cava (svc) and it frayed/ split when it was pulled into versacross connect sheath.There was resistance felt when withdrawing the wire from the introducer.Both the versacross connect and double curve sheath were removed from the body.Hence to resolve the issue, a new kit was opened and the procedure was completed successfully.No patient complications reported.The device is expected to be returned for analysis.
 
Additional Manufacturer Narrative
The device has not been received for analysis yet.However, based on the media provided, the reported allegations are confirmed.The photos clearly demonstrate the frayed mechanical guidewire.Also, the fields b7 (other relevant history) and e4 (init rptr also sent rep to fda) were updated.Thus, this supplemental is being filed.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.This product is part of the 97072033-fa epflex guidewire obstruction advisory commencement advisory for the j-tipped mechanical guidewire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACROSS CONNECT LAAC ACCESS SOLUTION
Common Device Name
DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
5825 explorer drive
mississauga, on L4W 5 P6
CA   L4W 5P6
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key17565394
Report Number2124215-2023-43649
Device Sequence Number17813364
Product Code DRE
UDI-Device Identifier00685447012580
UDI-Public00685447012580
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K220414
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/03/2025
Device Lot NumberVMFC290523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/26/2023
Supplement Date Received by Manufacturer02/15/2024
Initial Report FDA Received Date08/17/2023
Supplement Report FDA Received Date03/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age83 YR
Patient SexUnknown
-
-