| Catalog Number |
SCCS1002 |
| Medical Device Problem Codes |
Circuit Failure (1089); Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the sensica urine output device was checked in and no physical damage present.The device failed hi-pots test.
|
| |
|
Event or Problem Description
|
|
It was reported that the sensica urine output device was checked in and no physical damage present.The device failed hi-pots test.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported issue was confirmed.The root cause of the reported issue is a faulty battery.The device was evaluated upon receipt.The unit failed hi-pots test.The battery was replaced.A dhr is not required as this is not an out-of-box failure.The reported issue was confirmed through other elements of the investigation to not be labeling or packaging related.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
| |
|