• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. KIT INVISIKNOT ANKLE FRACTURE; WASHER, BOLT NUT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. KIT INVISIKNOT ANKLE FRACTURE; WASHER, BOLT NUT Back to Search Results
Catalog Number 72204834
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an ankle orif, the medial button of the kit invisiknot ankle was secured on the tibia and when the surgeon went to tighten down the suture the tightrope broke.The pieces were removed using tweezers.The procedure was completed with a s+n back up device.There was no surgical delay and no further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT INVISIKNOT ANKLE FRACTURE
Common Device Name
WASHER, BOLT NUT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17569459
Report Number1219602-2023-01530
Device Sequence Number8267479
Product Code HTN
UDI-Device Identifier00885556640227
UDI-Public00885556640227
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K162996
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72204834
Device Lot Number51082518
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2023
Supplement Date Received by Manufacturer11/06/2023
11/06/2023
Initial Report FDA Received Date08/17/2023
Supplement Report FDA Received Date11/09/2023
11/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age38 YR
Patient SexMale
-
-