Model Number 2244-01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bradycardia (1751); Non specific EKG/ECG Changes (1817); Hyperkalemia (4824)
Date of Event 07/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Block b3: approximated based on the year the article was published.Block e1: initial reporter facility name: (b)(6).Block d4, h4: the suspect device lot number was not provided in the article; therefore, the lot expiration and device manufacture dates are unknown.Block g2 literature source: inoue t, et al."hyperkalemia after endobiliary radiofrequency ablation for malignant biliary obstruction".Dig endosc.2021 jul;33(5):874-875.Doi: 10.1111/den.13989.Block h6: patient code (b)(6) captures the bradycardia.Patient code (b)(6) captures the hyperkalemia.Patient code (b)(6) captures the flattening p-waves and prolonged qrs complex.Impact code f2303 captures the calcium and glucose insulin was administered.
 
Event or Problem Description
Boston scientific became aware of an event involving a habib endohpb radio frequency ablation catheter through the article "hyperkalemia after endobiliary radiofrequency ablation for malignant biliary obstruction", by inoue t, et al.According to the literature, endoscopic transpapillary drainage failed because of tumor invasion to the duodenum, and endoscopic ultrasound-guided drainage was attempted.After penetrating the left intrahepatic duct via the stomach, a guidewire was passed through the lower bile duct stricture.The fistula was dilated using a 4-mm dilatation balloon catheter, and a habib endohpb radio frequency ablation catheter was inserted.Subsequently, the stricture was ablated at a power of 7 watts for 90 seconds.A metallic stent was deployed antegrade across the stricture, and a single-pigtail plastic stent was deployed across the fistula.The patients vital signs were stable during the procedure.However, severe bradycardia occurred 2 hours later with their heart rate plummeting below 20 to 30 beats per minute (bpm) and electrocardiogram showing flattened p-waves and prolonged qrs complex.A blood test revealed serum potassium levels of 7.4 meq/l, which had been 4.6 meq/l at pre-procedure.Calcium was administered immediately, and glucose-insulin therapy was initiated.One hour later, serum potassium levels decreased to 5.8 meq/l and stabilized at 4.8 meq/l 12 h later.The patients symptoms gradually improved.No further information has been obtained despite good faith efforts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HABIB ENDOHPB
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
STATICE MANUFACTURING
9 rue thomas edison
beacon 25000
FR   25000
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17571246
Report Number3005099803-2023-04371
Device Sequence Number15675043
Product Code KNS
UDI-Device Identifier08714729981909
UDI-Public08714729981909
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K180165
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Literature,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number2244-01
Device Catalogue Number2244-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/24/2023
Initial Report FDA Received Date08/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexMale
-
-