Catalog Number 8065992645
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Code Corneal Edema (1791)
Date of Event 05/02/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that metal particles were appeared at the wound during cataract surgery while using ophthalmic cannula.Procedure was completed on the same day.Additional information received indicating that the metal particles were observed at the main wound was retained.There was swelling in the wounds.The particles were not able to see in post operation due to edema.The patient was advised for a day magnetic resonance due to the material seems metallic.There was no intervention and hospitalization following the event.Ophthalmic knives were also used during procedure.
 
Additional Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of metal particles noted at the wound and swelling wound and corneal edema; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the manufacturing site and no lot information was provided; therefore, the root cause for the customer complaint issue cannot be determined.The cause of the reported event cannot be determined with the information obtained; therefore, specific action with regards to this complaint cannot be taken.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARCUT KNIVES
Common Device Name
KNIFE, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17571319
Report Number2523835-2023-00428
Device Sequence Number12817831
Product Code HNN
UDI-Device Identifier00380659926457
UDI-Public00380659926457
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065992645
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/25/2023
Supplement Date Received by Manufacturer10/15/2023
Initial Report FDA Received Date08/18/2023
Supplement Report FDA Received Date11/06/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CANNULA OPHTHALMIC; CUSTOM-PAK
Outcome Attributed to Adverse Event Other;
Patient Age68 YR
Patient SexFemale
-
-