| Catalog Number |
IR2641P |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the impaction of the 1st anchor, the roi-a cage got broken.There was a 20 minute delay to the procedure to install a new implant, and no reported patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed that the implant was fractured.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to off-axis forces applied on the cage during surgery or misalignment of the anchor in the cage.Dhr review: per dhr review, the part was likely conforming when it left zimvie control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that during the impaction of the 1st anchor, the roi-a cage got broken.There was a 20 minute delay to the procedure to install a new implant, and no reported patient impact.
|
| |
|